Features
· Only FDA approved natural
Kynar® or Polypropylene comes into contact with the process fluids
· Simple open construction is easily cleaned; manually, in place, or by steam
· Basic general purpose sanitary applications
· Suitable for the pharmaceutical industry
· Ladish Tri-Clamp® flange has a 316 S.S. backing for strength and maximum reliability
· Metal clamps hold full pressure seal even though all wetted materials are
plastic
· Suitable for use on most slurry and entrained solids applications
· 3A authorized and listed sanitary sensor
· Used in dairies and food processing plants
· A good choice for many cosmetic and printing plants
· Works with conductive or nonconductive liquids
Application
The Delta Controls Model P61 sensing probe is designed for sanitary service which requires superior cleanliness and high reliability. The wetted surfaces are composed entirely of virgin natural Kynar® (PVDF) or Polypropylene plastic. These surfaces have been welded together into a single piece without joints, crevices, or cracks. The finished unit is flame polished for the utmost in smoothness and cleanability.
The plastic flange facing has a 316 S.S. backing plate for strength and reliability. The basic sanitary flange mates with a Ladish Tri-Clamp® tubing size flange. Other type connections, such as ACME nut, APC, and Cherry-Burrell are optional.
Specifications
Working Pressure:
Up to 500 PSIG (35 BAR)
Working Temperature: -40 to +300°F (-40 to +150°C)
Insertion Length: Up to 8 feet (3 meters)
Process Connection: 1 inch or larger Tri-Clamp flange; other styles
and types optional
Electronic Module: Compatible with all Delta Controls Models
Process Temperature Sensing: Optional -125 to +250°F (-80 to
+122°C); 4-20 mADC output signal over a second 4-20 mA loop
Wetted Materials: Kynar® (PVDF) or Polypropylene
Body Materials: 316 Stainless Steel
Go to Delta Controls Home Page